Reproducibility of parafoveal choroidal vessel layer thickness measurements with enhanced depth imaging by spectral-domain optical coherence tomography 

ZHAO Jing, WANG Ya-xing, ZHANG Qi, LI Jian-jun, XU Liang.   

  1. Beijing Institute of Ophthalmology, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Key Laboratory of Ophthalmology and Visual Science, Beijing 100005, China
  • Received:2015-12-17 Online:2016-01-25 Published:2016-01-28
  • Contact: XU Liang, Email: xlbio1@163.com

摘要:

目的 研究采用频域相干光断层扫描(spectral-domain optical coherence tomography,SD-OCT)深度增强成像技术测量黄斑区脉络膜血管层厚度的观察者内以及观察者间的一致性。设计 前瞻性研究。研究对象 “北京眼病研究2011”可以测量脉络膜厚度的3187名受试者(平均年龄 64.0±9.6岁)中随机抽取的50人。方法 受试者接受SD-OCT检查。由两名培训过的眼科医生分别测量研究对象右眼黄斑中心凹、中心凹鼻侧750 μm及颞侧750 μm的脉络膜中小血管层和大血管层的厚度。采用组内相关系数(intra-class correlation coefficient, ICC)及Bland-Altman 法进行组内及组间一致性分析。主要指标 组内相关系数(ICC)及Bland-Altman表。结果 观察者组内ICC值均>0.945,Bland-Altman 法分析少于10.0% 的点在 95% 的一致性界限之外(P<0.05);观察者组间ICC值均>0.977,Bland-Altman 法分析少于8.0% 的点在 95% 的一致性界限之外(P<0.05)。结论 在常规测量条件下,SD-OCT测量脉络膜血管层厚度的观察者组内以及组间的一致性好,可辅助用于眼底疾病的临床诊断研究。(眼科,2016, 25: 51-57)

关键词: 脉络膜厚度, 一致性分析

Abstract:

Objective To measure the inter-observer reproducibility and intra-observer reproducibility of choroidal vessel layer thickness measurements of macula measured with spectral-domain optical coherence tomography(SD-OCT). Design Prospective study. Participants 50 individuals selected randomly from the Beijing Eye Study 2011, which included 3187 individuals with a mean age of 64.0±9.6 years that choroidal vessel layer thickness can be measured. Methods The medium choroidal vessel-choriocapillaris layer and the large choroidal vessel layer were measured by two ophthalmologists with enhanced depth imaging of SD-OCT. Only the right eye of each participant was assessed and the measurements were performed at the following locations:  the fovea, 750 μm nasal to the fovea and 750 μm temporal to the fovea. The intra-class correlation coefficient (ICC) and Bland-Altman analysis were calculated. Main Outcome Measures The intra-class correlation coefficient (ICC) and Bland-Altman plot. Results The analysis of the intra-observer variability revealed an intra-class correlation coefficient( ICC) of more than 0.945 (P<0.05), thereafter, Bland-Altman plot showed less than 10.0% points outside the 95% limits of agreement. We also found an inter-observer ICC reproducibility value of more than 0.977, whereupon, Bland-Altman plot showed less than 8.0% points outside the 95% limits of agreement. Conclusion Under routine examination conditions, measurements of choroidal vessel layer thickness by SD-OCT showed almost perfect intra-observer reproducibility and inter-observer reproducibility, this procedure is suitable for clinical diagnosis and research of retinal disease.  (Ophthalmol CHN, 2016, 25: 51-57)

Key words: choroidal thickness, reproducibility